Lila is the assistant nursing supervisor of the intensive care unit at Matthew J. Ryan Veterinary Hospital of University of Pennsylvania. She has lectured at state conferences and has published chapters in various veterinary manuals. Her passion lies in all aspects of advanced nursing care, including patient advocacy, innovative nursing development, and education of prospective veterinarians and nurses.
Read Articles Written by Lila K. SierraEmily Savino
BA, CVT, VTS (Emergency and Critical Care) | Matthew J. Ryan Veterinary Hospital, University of Pennsylvania
Emily is the nursing supervisor of the intensive care unit at Matthew J. Ryan Veterinary Hospital of University of Pennsylvania. She has lectured on veterinary critical care nursing at conferences, and has published chapters in veterinary texts. She enjoys teaching new veterinary nurses the art of “critical thinking” and working in an environment where she is learning new things every day.
Read Articles Written by Emily Savino
Systemic arterial blood pressure is created by the pumping action of the heart, circulating arterial blood volume, and the smooth muscle tone of blood vessel walls.
Arterial blood pressure is essential for adequate perfusion of tissues, delivering oxygen for energy demands. The ability to measure and monitor blood pressure trends provides important data about patient cardiovascular status and may help define approach to treatment.
This article was originally published in the March/April 2015 “Today’s Technician” column of Today’s Veterinary Practice. Today’s Veterinary Practice is the NAVC’s official, peer-reviewed journal for veterinarians.
Part 1 of this article series—Overview of Blood Pressure Monitoring (January/February 2015 issue)—discussed terms associated with blood pressure measurement, indications for measurement, and normal and abnormal blood pressure values. This second article describes types of blood pressure monitors and provides step-by-step instructions on performing blood pressure measurement.
MONITORING TECHNIQUES
Arterial blood pressure can be measured in 2 ways:
- Direct arterial blood pressure (DABP) monitoring—considered the gold standard—uses an arterial catheter connected to a pressure transducer. This system allows continuous monitoring of patient systolic, diastolic, and mean arterial pressure (SAP, DAP, and MAP, respectively)1-3 and also simplifies collection of samples for arterial blood gas analysis. However, DABP monitoring is not used as frequently as other methods because an arterial catheter must be placed.
- Indirect arterial blood pressure (IABP) monitoring relies on noninvasive detection of arterial blood flow or vessel wall movement in a peripheral artery and provides intermittent measurements of arterial blood pressure.IABP monitoring is the most frequently used technique in clinical practice; the most commonly used indirect methods are Doppler ultrasonography and oscillometric blood pressure monitoring. To obtain an IABP reading, a cuff is inflated over an artery until arterial blood flow is occluded.1
Normal arterial blood pressure values for adult dogs and cats are listed in Table 1.
TABLE 1. Normal Arterial Blood Pressure Values in Adult Dogs & Cats4 |
||
BLOOD PRESSURE VALUES | DOGS | CATS |
Systolic arterial pressure | 90 to 140 mm Hg | 80 to 140 mm Hg |
Diastolic arterial pressure | 50 to 80 mm Hg | 55 to 75 mm Hg |
Mean arterial pressure | 60 to 100 mm Hg | 60 to 100 mm Hg |
INDIRECT BLOOD PRESSURE MONITORING:
Doppler Ultrasonography
Doppler technology uses 10 MHz ultrasound waves to detect blood flow in a peripheral artery (Figure 1), which is then made audible via the probe/crystal and a speaker.1
The Doppler technique does not provide a measurement of MAP or DAP in small animals.1,4,5 While some studies have reported that Doppler readings may more closely reflect MAP than SAP in cats,1,3 the most recent study was performed on healthy anesthetized cats.1 Therefore, most clinicians consider that Doppler readings in awake animals estimate SAP.

FIGURE 1. Doppler ultrasonographic blood pressure measurement.
Advantages
Doppler ultrasonography has several advantages when compared with other blood pressure monitoring techniques; the technology is:
- Relatively easy to use, not requiring a great deal of expertise
- Affordable and readily available in most clinical settings
- Appropriate to use in patients with significant hypotension and cardiac arrhythmias1
- Commonly used in small animals (cats, small dogs, rabbits, ferrets, other exotic species).
Blood Pressure Measurement: Step by Step
Table 2 lists the supplies needed for Doppler ultrasonographic blood pressure measurement.
TABLE 2. Required Supplies for Doppler Measurement |
|
- Select a peripheral artery for Doppler probe placement; common locations include the dorsal pedal artery, digital artery, and coccygeal artery.
- Clip the hair coat over the chosen artery, or wet it with alcohol, and apply ultrasound coupling gel to the Doppler probe.
- Place the Doppler prober over the artery, oriented parallel to blood flow (Figure 2); the probe is considered correctly placed when pulsatile blood flow is audible.
- Connect a pressure cuff to a sphygmomanometer and place the cuff proximal to the Doppler probe.
- Inflate the cuff until the artery is occluded and Doppler sounds are no longer audible; then slowly deflate the cuff, while observing the sphygmomanometer.
- Record the pressure at which the first audible arterial pulse is heard as the SAP.
- Obtain several consecutive measurements; average the middle 3 readings, and record the average as the SAP. Make sure to use the same limb and same size cuff for serial blood pressure measurements.
Preventing Inaccurate Results
Several factors can result in erroneous readings:
- Cuff size is critically important to obtain accurate Doppler readings.1,4,6
- A cuff that is too large falsely decreases the reading
- A cuff that is too small falsely elevates the reading.
- Patients must be held/restrained during Doppler measurements; however, this handling may increase the animal’s blood pressure.
- Patient position during Doppler measurement can contribute to inaccurate readings; ideally, patients should lie in lateral recumbency, with the cuff positioned on the limb at the level of the right atrium.1,4
- Obtaining Doppler measurements in patients with severe peripheral vasoconstriction (hypovolemia, shock, significant hypothermia) can be difficult; therefore, accuracy of readings in these patients can be affected by degree of vasoconstriction.3,4
INDIRECT BLOOD PRESSURE MONITORING: Oscillometric Measurement
Oscillometric measurement (Figure 3) provides values for SAP, MAP, DAP, and pulse rate; therefore, this type of blood pressure monitoring provides more information than Doppler measurement.1,3,4 Most oscillometric devices measure the MAP; then calculate the SAP and DAP via programmed algorithms; therefore, the MAP is the most reliable reading.4

FIGURE 3. Oscillometric blood pressure monitor (Cardell 9401 Veterinary Monitor, midmark.com).
Advantages
Oscillometric blood pressure measurement has several advantages over Doppler methodology; the:
- Process is more automated, requiring less technical skill; the operator simply chooses the appropriate cuff size, places it on the patient, and hits the start button
- Monitors can be programmed to measure blood pressure at timed intervals (eg, Q 15 min)
- Automated nature of measurement helps ensure that elevated values are less likely a result of stressful patient handling.1,4
Disadvantages
However, oscillometric techniques tend to be less accurate in patients:3,4
- Under 5 kg
- With cardiac arrhythmias, significant tachycardia or bradycardia, vasoconstriction, or hypothermia
- That are moving or shivering.
Blood Pressure Measurement: Step by Step
- Place the cuff—attached to an oscillometric monitor—on a distal artery; common locations for oscillometric cuff placement are similar to those used for Doppler blood pressure measurement.
- Once the device is started, the cuff automatically inflates to a pressure that occludes arterial blood flow.
- As the cuff is deflated, the arterial wall oscillations increase at SAP, maximize at MAP, and decrease at DAP.4
- The oscillometric monitor display will show numeric values for SAP, MAP, and DAP as well as pulse rate.
- Compare the oscillometric pulse rate with a manually obtained pulse rate to determine accuracy: if the rates do NOT match, it is likely that the blood pressure readings on the monitor are inaccurate.
Preventing Inaccurate Results
As discussed in the Indirect Blood Pressure Monitoring: Doppler Ultrasonography section:
- Selection of an appropriately sized cuff is crucial to obtaining accurate values; cuff selection is outlined in Table 2.
- To ensure accuracy, it is important to take several consecutive readings; then compare the readings to ensure that the values are reproducible.
- Ideally, the same size cuff and same limb should be used for all consecutive blood pressure measurements.
DIRECT BLOOD PRESSURE MONITORING
In general, DABP monitoring (Figures 4 and 5) is indicated for any critical patient, but specific indications in the clinical setting include patients:
- Presenting in hypovolemic or septic shock
- In congestive heart failure, especially whenreceiving powerful vasodilator medication forpurpose of afterload reduction
- Requiring vasopressors or mechanical ventilation
- Receiving medication for severe hypertension
- Demonstrating a high anesthetic risk.
DABP monitoring is NOT indicated in healthy, ambulatory patients because these patients are more likely to disconnect an arterial line, or remove the arterial catheter, increasing the risk for arterial hemorrhage.

FIGURE 4. Simultaneous electrocardiography and direct arterial blood pressure readings (Escort Prism Patient Monitor, invivocorp.com)

FIGURE 5. Simultaneous electrocardiography and direct arterial blood pressure readings demonstrated on a different monitor (PM-9000 Vet Monitor, mindray.com).
Advantages
DABP has many advantages over other methods of blood pressure monitoring. Two of the most important advantages include:
- “Real-time” monitoring of blood pressureand blood pressure trends, even in extremely hypotensive patients; this up-to-the-minute information about a patient’s hemodynamic status enables clinicians to gauge whether a specific therapy is working or if additional, immediate intervention is necessary.
- Handling and restraint of the patient is usually limited to the initial arterial catheter placement; therefore, DABP values are less prone to falsely elevated readings related to stress of restraint and handling.
Disadvantages
However, DABP monitoring is not used extensively because it has several drawbacks:
- The equipment necessary to monitor DABP (eg,pressure transducers, hemodynamic monitors) can be cost prohibitive, especially when not used frequently.
- Arterial catheters are invasive, and arterial access can be technically difficult to obtain and maintain.
- Complications that can result from arterial catheterization include bleeding from the catheter insertion site, hematoma formation, significant hemorrhage if the system becomes disconnected, infection, and arterial thrombosis with possibility of necrosis of the tissues distal to the catheter.
Blood Pressure Measurement: Step by Step
Table 3 lists the supplies needed for DABP measurement (Figure 6).
TABLE 3. Required Supplies for DABP Measurement |
|

FIGURE 6. DABP supplies: Transpac disposable pressure transducer (icumed.com), heparinized saline bag inside pressurized air bag, electronic cable to connect Transpac to monitor, transducer board (white), arterial catheter (dorsal metatarsal).
- Place and secure an arterial catheter; common sites for arterial catheterization include the dorsal pedal artery, coccygeal artery (Figures 7 and 8), and medial auricular artery.1,2
- Once the arterial catheter is placed and secured, connect it to the DABP monitoring system.
- At one end of the system—the end farthest away from the patient—a fluid administration set is connected to a heparinized bag of 0.9% sodium chloride (Table 3).
- Pressurize the fluid bag to a range between 250 and 300 mm Hg: the goal is to achieve a pressure in the fluid bag greater than the patient’s systolic pressure, which prevents back flow of arterial blood into the monitoring system.1,2
- The fluid bag tubing connects to a pressure transducer, which is connected by a cable to the physiologic monitor, mounted on a board placed at the level of the patient’s heart. At the other end of the transducer, semi-rigid tubing connects to the arterial catheter’s T-set.
- Once the pressure transducer is connected to the monitor, zero it at the level of the patient’s right atrium. Once zeroed, the pressure transducer converts the pressure changes in the artery to an electrical signal that is displayed on the monitor as a pressure wave form; numeric values for SAP, MAP, and DAP are also displayed.1
- Flush the DABP monitoring system with heparinized saline before connecting it to the patient.
- Supervise patients with arterial catheters at all times and carefully care for the catheter:
- Clearly label the catheter, allowing easyidentification as arterial rather than venous and, thereby, alerting personnel NOT to administer medications through the catheter.
- Rewrap the catheter, at a minimum, once daily; rewrap soiled catheters promptly.
- While the catheter is unwrapped, evaluate the insertion site for redness, warmth, swelling, pain, or discharge; remove catheters that are painful or oozing from the insertion site.
- Document when the catheter is rewrapped and note the appearance of the insertion site in the medical record.
- Consider putting an E-collar on patients to prevent access to arterial catheters.
- Do NOT maintain arterial catheters in cats for longer than 6 to 12 hours1 owing to increased risk for arterial thrombosis.
Preventing Inaccurate Results
Although DABP is considered the gold standard in blood pressure monitoring, a number of situations can cause erroneous readings, which can be prevented by:
- Using only semi-rigid tubing to connect the pressure transducer to the patient—the use of compliant tubing results in excessive damping of the pressure wave, causing inaccurate readings
- Ensuring air bubbles are not present in the system
- Checking the semi-rigid tubing for kinks
- Making sure the pressure bag is inflated to at least 250 mm Hg
- Reassessing the transducer level; if the transducer is displaced, replace it to approximately the level of the right atrium and zero to the patient
- Ensuring that the arterial catheter is patent; flush the arterial catheter if needed.
IN SUMMARY
Veterinary nurses play a fundamental role in patient care. Appropriately monitoring patient blood pressure, documenting results, interpreting them, and communicating changes and concerns to the clinician provides all members of the veterinary team with an opportunity to proactively manage potential problems.
DABP = direct arterial blood pressure; DAP = diastolic arterial pressure; IABP = indirect arterial blood pressure; MAP = mean arterial pressure; SAP = systolic arterial pressure
References
- Waddell LS, Brown AJ. Hemodynamic monitoring. In Silverstein DC, Hopper K (eds): Small Animal Critical Care Medicine, 2nd ed. St. Louis: Elsevier, 2015, pp 957-962.
- Cooper E, Cooper S. Direct systemic arterial blood pressure monitoring. In Burkitt Creedon JM, Davis H (eds): Advanced Monitoring and Procedures for Small Animal Emergency and Critical Care. Ames, IA: Wiley-Blackwell, 2012, pp 122-133.
- Cooper E. Hypotension. In Silverstein DC, Hopper K (eds): Small Animal Critical Care Medicine, 2nd ed. St. Louis: Elsevier, 2015, pp 46-50.
- Williamson JA, Leone S. Noninvasive arterial blood pressure monitoring. In Burkitt Creedon JM, Davis H (eds): Advanced Monitoring and Procedures for Small Animal Emergency and Critical Care, Ames, IA: Wiley-Blackwell, 2012, pp 134-144.
- Caulkett NA, Cantwell SL, Houston DM. A comparison of indirect blood pressure monitoring techniques in the anesthetized cat. Vet Surg 1998; 27:370.
- Monnet E. Cardiovascular monitoring. In Wingfield WE, Raffe MR (eds): The Veterinary ICU Book. Jackson, WY: Teton NewMedia, 2002, pp 265-280.